4.              WATER QUALITY

4.1                Monitoring Requirement

4.1.1                In accordance with the EM&A Manual, impact monitoring is conducted for three days per week at mid-flood and mid-ebb with sampling and measurement at the designated monitoring stations.

4.2                Monitoring Equipment

4.2.1                Equipment used for in-situ measurement and water sampling during impact water quality monitoring is summarised in Table 12. The equipment is in compliance with the requirements set out in the EM&A Manual. All in-situ monitoring instruments were calibrated by a HOKLAS- accredited laboratory. Calibration of temperature, DO, salinity, pH and turbidity is conducted in three-month interval. Calibration certificates for the water quality monitoring equipment are attached in Appendix D.

4.3                Equipment Calibration

4.3.1                All in-situ monitoring instruments shall be checked, calibrated and certified by a laboratory accredited under HOKLAS before use and subsequently re-calibrated at three monthly intervals throughout all stages of the water quality monitoring programme. Responses of sensors and electrodes shall be checked with certified standard solutions before each use. Wet bulb calibration for a DO meter shall be carried out before measurement at each monitoring location.

4.3.2                Sufficient stocks of spare parts shall be maintained for replacements when necessary. Backup monitoring equipment shall also be made available so that monitoring is uninterrupted even when some equipment is under maintenance or calibration etc.

4.4                Monitoring Parameters

The monitoring parameters and frequency for both in-situ measurement and laboratory analysis are summarised in Table 13.

4.5                Monitoring Operation

4.5.1                The position of water monitoring station will be located by the Differential Global Positioning System (DGPS) or equivalent. The water depth of water monitoring station will be determined by the echo sounder affixed to the bottom of the monitoring vessel or a portable echo sounder depth detector.

4.5.2                Once the location and water depth are confirmed, water samples shall be collected at 3 depths (1m below the surface, mid-depth, and 1m above the seabed) of the water column at each location, except where water depth is less than 6m, the mid-depth will be omitted and if the water depth is less than 3m only the mid-depth station will be monitored. Duplicate marine samples will be collected in each sampling event. The water samples are decanted from the water sampler into the water sample bottles. The bottles are labelled, tightly sealed, placed into a cool-box and packed with ice ready for delivery to the laboratory.

4.5.3                Two consecutive measurements of water quality data, including pH, salinity, dissolved oxygen and turbidity will be recorded according to the monitoring locations. Separate deployment of the monitoring instruments and water samplers will be conducted for the consecutive measurements or samplings. The monitoring location / position, time, water depth, sampling depth, tidal stages, weather conditions, sea condition and any special phenomena or work underway nearby shall also be recorded. If the difference in value between the first and second measurement of DO or turbidity parameters is more than 25% of the value of the first reading, the reading shall be discarded and further readings should be taken.

4.6                Laboratory Measurement / Analysis

Background

4.6.1                Acumen Laboratory and Testing Limited (HOKLAS Reg: No.241) has been appointed to conduct the laboratory measurement or analysis of water sample in this project.

Quality Assurance / Quality Control

4.6.2                The laboratory incorporates a variety of QA/QC monitoring programme into their testing system. Where applicable or available, the quality of the analysis will be monitored by conducting the following QC analysis:

For each batch of 20 samples:

·           A minimal of 1 laboratory method blank will be analyzed;

·           A minimal of 1 sample duplicate will be analyzed;

·           A minimal of 1 sample matrix spike will be analyzed.

4.7                Monitoring Locations

4.7.1                In accordance with the EM&A Manual, water quality monitoring should be carried out at 3 designated monitoring locations.

4.7.2                The coordinates of the monitoring location stated in the EM&A Manual is summarised in Table 14 and the locations of the water quality monitoring stations shown in Figure 4.

4.8                Monitoring Results

4.8.1                The schedule of water quality monitoring in reporting month is provided in Appendix E.

4.8.2                Impact water quality monitoring was conducted at all designated monitoring stations in the reporting month. Impact water quality monitoring results and graphical presentations are provided in Appendix F.

4.8.3                Typhoon Signal No. 8 was hoisted on 1 September 2023. Due to safety concerns, the water quality monitoring on 1 September 2023 has been cancelled.

4.8.4                Black Rainstorm Signal was hoisted on 8 September 2023. Due to safety concerns, the water quality monitoring on 8 September 2023 has been cancelled.

4.8.5                The weather and meteorological conditions during the monitoring are provided in Appendix K.

4.8.6                Number of Action/ Limit exceedance recorded in the reporting month at each impact stations is summarized in Table 15.

4.8.7                During the reporting period, no Action and Limit Level exceedance was recorded for water quality monitoring.

4.8.8                The Event and Action Plan for water quality is given in Appendix H.

4.9                WetSeps

Three WetSeps are deployed within the site for treatment of the site runoff prior to disposal in compliance with the conditions stipulated in the water discharge license (Variation of WPCO Discharge Licence was approved by EPD on 1 December 2022 with immediate effect).