4.                   Water Quality Monitoring

4.1                   Monitoring Requirement

4.1.1                                 With reference to Section 4.6.2 of the EM&A manual, the baseline water quality shall be carried out once per day for 3 days per week, for a period of 4 weeks prior to the commencement of construction works. The interval between two sets of monitoring shall not be less than 36 hours.

4.1.2                                 There shall not be any major construction activities in the vicinity of the stations during the baseline monitoring.

4.2                   Monitoring Locations

4.2.1                                 In accordance with Section 4 of the EM&A Manual, water quality monitoring should be carried out at 11 designated monitoring locations (3 stations for construction phase and 8 stations for operation phase).

4.2.2                                 A proposal of relocation of water quality monitoring locations was submitted to EPD on 22 September 2020 in accordance with Section 4.5.7 of the EM&A Manual, the monitoring location E2, E3 and E5 are proposed to be relocated by alternative monitoring location E2a, E3a and E5a. EPD¡¦s approval letter is given in Appendix I.

4.2.3                                 The water quality monitoring locations are shown in Figure 4.

4.2.4                                 The coordinates of the monitoring location stated in the EM&A Manual is summarised in Table 4.1.

4.3                   Monitoring Parameters

4.3.1                                 Parameters for each monitoring station for both in-situ measurement and laboratory analysis are summarised in Table 4.2. 

4.4                   Monitoring Equipment

4.4.1                                 The following equipment and facilities shall be used for the monitoring of water quality impacts:

Dissolved Oxygen and Temperature Measuring Equipment

4.4.1.1            The instrument shall be a portable and weatherproof DO measuring instrument complete with cable and sensor, and use a DC power source. The equipment shall be capable of measuring:

¡P         a DO level in the range of 0 ¡V 20 mg L-1 and 0 ¡V 200% saturation; and

¡P         a temperature of 0 ‑ 45 degree Celsius.

4.4.1.2            It shall have a membrane electrode with automatic temperature compensation complete with a cable. Sufficient stocks of spare electrodes and cables shall be available for replacement where necessary.

4.4.1.3            Shall salinity compensation not be built-in to the DO equipment, in-situ salinity shall be measured to calibrate the DO equipment prior to each DO measurement.

Turbidity Measurement Instrument

4.4.1.4            Turbidity shall be measured in-situ by the nephelometric method. The instrument shall be portable and weatherproof turbidity measuring instrument using a DC power source complete with cable, sensor and comprehensive operation manuals. It shall have a photoelectric sensor capable of measuring turbidity between 0 - 1000 NTU (for example, Hach model 2100P or an approved similar instrument). The cable shall not be less than 25m in length. The meter shall be calibrated to establish the relationship between NTU units and the levels of suspended solids. The correlation between Turbidity (NTU) & Suspended Solids (mg/L) which was established in the baseline monitoring under the project are shown in Appendix A.

Sampler

4.4.1.5            A water sampler is required. It shall comprise a transparent Polyvinyl Chloride (PVC) cylinder, with a capacity of not less than 2 liters, which can be effectively sealed with latex cups at both ends. The sampler shall have a positive latching system to keep it open and prevent premature closure until released by a messenger when the sampler is at the selected water depth (for example, Kahlsico Water Sampler or an approved similar instrument).

Water Depth Detector

4.4.1.6            A portable, battery-operated echo sounder shall be used for the determination of water depth at each designated monitoring station. This unit can either be hand held or affixed to the bottom of the work boat, if the same vessel is to be used throughout the monitoring programme.

Salinity

4.4.1.7            A portable salinometer capable of measuring salinity in the range of 0 - 40 parts per thousand (ppt) shall be provided for measuring salinity of the water at each monitoring location.

pH

4.4.1.8            The instrument shall consist of a potentiometer, a glass electrode, a reference electrode and a temperature-compensating device. It shall be readable to 0.1 pH in a range of 0 to 14. Standard buffer solutions of at least pH 7 and pH 10 shall be used for calibration of the instrument before and after use. Details of the method shall comply with American Public Health Association (APHA), 19th ed. 4500-HTB.

Sample Containers and Storage

4.4.1.9            Water samples shall be stored in high density polythene bottles with no preservative added, packed in ice (cooled to 4¢XC without being frozen) and delivered to the laboratory and analyzed within 24 hours after collection. Sufficient volume of samples shall be collected to achieve the required detection limit.

The container types for holding water samples are presented in Table 4.3.

4.4.1.10        Monitoring Position Equipment

A hand-held or boat-fixed type digital Differential Global Positioning System (DGPS) with way point bearing indication or other equipment instrument of similar accuracy, shall be provided and used during marine water monitoring to ensure the monitoring vessel is at the correct location before taking measurements.

4.4.1.11        Current Velocity and Direction

Valeport Model 106 is proposed for monitoring the current velocity and direction except at M3 where the vessel cannot be accessed and River Surveyor M9 is proposed to be used for current velocity and direction.

4.4.2                                 The equipment employed for the monitoring and sampling and their specifications are presented in Table 4.4.

4.5                   Equipment Calibration

4.5.1                                 All in-situ monitoring instruments shall be checked, calibrated and certified by a laboratory accredited under HOKLAS before use and subsequently re-calibrated at three monthly intervals throughout all stages of the water quality monitoring programme. Responses of sensors and electrodes shall be checked with certified standard solutions before each use. Wet bulb calibration for a DO meter shall be carried out before measurement at each monitoring location.

4.5.2                                 Sufficient stocks of spare parts shall be maintained for replacements when necessary. Backup monitoring equipment shall also be made available so that monitoring is uninterrupted even when some equipment is under maintenance or calibration etc.

4.6                   Monitoring Operation

4.6.1                                 The position of water monitoring station will be located by the Differential Global Positioning System (DGPS) or equivalent. The water depth of water monitoring station will be determined by the echo sounder affixed to the bottom of the monitoring vessel or a portable echo sounder depth detector.

4.6.2                                 Once the location and water depth are confirmed, water samples shall be collected at 3 depths (1m below the surface, mid-depth, and 1m above the seabed) of the water column at each location, except where water depth is less than 6m, the mid-depth will be omitted and if the water depth is less than 3m only the mid-depth station will be monitored. Duplicate marine samples will be collected in each sampling event. The water samples are decanted from the water sampler into the water sample bottles. The bottles are labelled, tightly sealed, placed into a cool-box and packed with ice ready for delivery to the laboratory.

4.6.3                                 Two consecutive measurements of water quality data, including pH, salinity, dissolved oxygen and turbidity will be recorded according to the monitoring locations. Separate deployment of the monitoring instruments and water samplers will be conducted for the consecutive measurements or samplings. The monitoring location / position, time, water depth, sampling depth, tidal stages, weather conditions, sea condition and any special phenomena or work underway nearby shall also be recorded. If the difference in value between the first and second measurement of DO or turbidity parameters is more than 25% of the value of the first reading, the reading shall be discarded and further readings should be taken.

4.7                   Laboratory Measurement / Analysis

Background

4.7.1                                 Fugro Technical Services Limited (HOKLAS Reg: No.015) has been appointed to conduct the laboratory measurement or analysis of water sample in this project. The certificate of accreditation are shown in Appendix J.

Quality Assurance / Quality Control

4.7.2                                 The laboratory incorporates a variety of QA/QC monitoring programme into their testing system. Where applicable or available, the quality of the analysis will be monitored by conducting the following QC analysis:

For each batch of 20 samples:

¡±  A minimal of 1 laboratory method blank will be analyzed;

¡±  A minimal of 1 sample duplicate will be analyzed; 

¡±  A minimal of 1 sample matrix spike will be analyzed.

4.7.3                                 QA/QC records is provided in Appendix D.

4.8                   Results and Observations

4.8.1                                 Baseline water quality monitoring was conducted at 11 designated monitoring locations between 1 December 2020 and 26 December 2020. The baseline monitoring schedule is given in Appendix F.

4.8.2                                 No major construction activities were observed during the period of baseline monitoring. The weather condition in the monitoring period is provided in Appendix G.

4.8.3                                 A summary of the in-situ baseline water quality monitoring results is given in Table 4.5. Results of laboratory analysis of baseline water quality are presented in Table 4.6. Detailed water quality monitoring results is provided in Appendix D.

4.9                   Action and Limit Level

4.9.1                                 The Action and Limit Levels in construction phase water quality monitoring were determined in accordance with the EM&A Manual, as shown in Table 4.7. Action and Limit Levels are used to determine whether operational modifications are necessary to mitigate impacts to water quality.

4.9.2                                 The Action Levels (AL) and Limit Levels (LL) of water quality are determined from the baseline monitoring data and are shown in Table 4.8.

4.10               Event and Action Plan

4.10.1                           The Event and Action Plan for Water Quality are given in Appendix E.